Nidek Tonoref IIHow do I clean the NIDEK TONOREF II air nozzle after a pressure error?To clean the NIDEK TONOREF II air nozzle, display the NT measuring screen, inspect the glass from an oblique angle, blow off…Read: How do I clean the NIDEK TONOREF II air nozzle after a pressure error?
Bausch & Lomb Stellaris Phaco MachineWhat does LM24 mean when the Bausch & Lomb Stellaris Phaco Machine says an illuminator bulb is at end of life?LM24 means the upper illuminator bulb is at end of life and will be disabled if turned off. Replace the upper illuminator bulb.Read: What does LM24 mean when the Bausch & Lomb Stellaris Phaco Machine says an illuminator bulb is at end of life?