Essilor Mr Blue 2.0 Edger and Tracer setHow often should I replace the drill bit on the Essilor Mr Blue 2.0?Replace the 0.8 mm drill bit every 250 holes and the 1.0 mm drill bit every 500 holes. Installing a new bit resets the wear…Read: How often should I replace the drill bit on the Essilor Mr Blue 2.0?
Essilor Mr Blue 2.0 Edger and Tracer setHow do I open the cleaning cycle on the Essilor Mr Blue 2.0?From the edger work screen, select the cleaning menu, then choose door cleaning or module-and-trough cleaning. Door cleaning…Read: How do I open the cleaning cycle on the Essilor Mr Blue 2.0?
Essilor Mr Blue 2.0 Edger and Tracer setWhat do the three small-tools wear thresholds mean on the Essilor Mr Blue 2.0?The first threshold shows a maintenance icon, the second adds the message “Maintenance is needed. Have you contacted your service…Read: What do the three small-tools wear thresholds mean on the Essilor Mr Blue 2.0?
Essilor Mr Blue 2.0 Edger and Tracer setWhen does the Essilor Mr Blue 2.0 ask for the mill bit to be changed?The mill-bit wear indicator is triggered after 500 machined lenses, but MHI lenses count 2.5 times faster. For example, 200 MHI…Read: When does the Essilor Mr Blue 2.0 ask for the mill bit to be changed?
Nidek Tonoref IIHow do I clean the NIDEK TONOREF II air nozzle after a pressure error?To clean the NIDEK TONOREF II air nozzle, display the NT measuring screen, inspect the glass from an oblique angle, blow off…Read: How do I clean the NIDEK TONOREF II air nozzle after a pressure error?
Bausch & Lomb Stellaris Phaco MachineWhat does LM24 mean when the Bausch & Lomb Stellaris Phaco Machine says an illuminator bulb is at end of life?LM24 means the upper illuminator bulb is at end of life and will be disabled if turned off. Replace the upper illuminator bulb.Read: What does LM24 mean when the Bausch & Lomb Stellaris Phaco Machine says an illuminator bulb is at end of life?